• Color/absorption measurement in liquids
  • Concentration measurement of substances by absorption at characteristic wavelengths
  • Calibration in E/m, APHA-Hazen, EBC, ASTM, Saybolt, etc.
  • Dual-beam measurement for high stability
  • Installation on standard Varivent® housing or custom-specific flow cells
  • Fast and simple verification with control unit
  • Option: Turbidity compensation using a second wavelength

The ColorPlus measures the colour/absorbance online at up to three different wavelengths in the UV and visible range, also combined. Turbidity can be compensated at the same time by selecting an appropriate wavelength. The modular design of the instrument enables to cover various applications by selecting the most suitable measuring cell. In many cases, a standard housing like e.g. Varivent® is used. Alternatively, custom-specific flow cells in different materials and designs are available. For inspecting the instrument, checking units on the basis of reference filters can easily be inserted. Operation is done intuitive using the intelligent control system SICON via touch screen technology and colour display. Values, graphs, alarm and status messages can be presented. An internal data logger allows recalling and displaying measured data from the last 32 days. A variety of analogue and digital interfaces allows smooth system integration and control.

Measuring principle

Absorption

Nominal range

0 .. 3 E

Scale ranges

8 ranges, freely configurable

Resolution

0.001 E

Flow cell material

Stainless steel 316L, Titanium, Hastelloy, PVDF, etc.

Ambient temperature

-20 .. +50 °C

Flow cells

Varivent®, custom-specific

Wavelengths UV

254, 313, 365, 436, 546 nm

Wavelengths VIS

365, 380 .. 700 nm

Units

E, E/m, APHA-Hazen, EBC, Saybolt, ASTM, GOST, etc.

Lowest measuring range 0-50 EBC TurBiScat / 0-5 EBC ColorPlus
Highest measuring range 0-50 EBC TurBiScat / 0-208 EBC ColorPlus
Applications in Brewing Industry
Colour (EBC) of Beer on the Blending Unit
Lowest measuring range 0-50 EBC TurBiScat / 0-5 EBC ColorPlus
Highest measuring range 0-50 EBC TurBiScat / 0-208 EBC ColorPlus
Applications in Brewing Industry
Colour (EBC) of Hot Wort before Cooler
Lowest measuring range 0-10 EBC
Highest measuring range 0-160 EBC (max. 30 EBC Trübung / Turbidity)
Applications in Brewing Industry
Colour (Hazen) of Maleic Anhydride
Lowest measuring range 0-0.05 E
Highest measuring range 0-3 E
Applications in Chemical / Pharmaceutical Industry
Petrochemical Industry / Refinery
Colour (ICUMSA) of Liquid Sugar
Lowest measuring range 0-0.05 E
Highest measuring range 0-3 E
Applications in Sugar Industry
Colour of Spirits at the Blending Unit
Lowest measuring range 0-5 EBC F
Highest measuring range 0-208 EBC F
Applications in Spirits Industry
Concentration of Chlorine Dioxide (ClO2)
Lowest measuring range 0-0.3 g/l
Highest measuring range 0-15 g/l
Applications in Chemical / Pharmaceutical Industry
Drinking Water Treatment
Pulp / Paper Industry
Concentration of Chlorine in Nitrosyl Chloride (NOCl)
Lowest measuring range 0-0.05 E
Highest measuring range 0-3 E
Applications in Chemical / Pharmaceutical Industry
Concentration of Coffee in Condensate
Lowest measuring range 0- 0.05 E
Highest measuring range 0-3 E
Applications in Food Industry
Concentration of Copper in Sulphuric Acid (H2SO4)
Lowest measuring range 0-0.05 E
Highest measuring range 0-3 E
Applications in Chemical / Pharmaceutical Industry
Machine and Metal Industry
Petrochemical Industry / Refinery
Concentration of Flavonoids in Lemon Juice
Highest measuring range 0-900 ppm
Applications in Food Industry
Concentration of Hypochlorite (ClO-) in Water
Lowest measuring range 0-2.1 mg/l
Highest measuring range 0-6.3 g/l
Applications in Chemical / Pharmaceutical Industry
Concentration of Rare-Earth
Lowest measuring range 0-10mg/l
Highest measuring range 0-1g/l
Applications in Chemical / Pharmaceutical Industry
Concentration of SurTec 650 (chromitAL) in Demineralized Water
Lowest measuring range 0-0.1 E
Highest measuring range 0-3 E
Applications in Machine and Metal Industry
DOC (UV Absorption) after Activated Carbon Filtration
Lowest measuring range 0-1 E/m
Highest measuring range 0-60 E/m
Applications in Drinking Water Treatment
DOC (UV Absorption) in Raw Water
Lowest measuring range 0-1 E/m
Highest measuring range 0-60 E/m
Applications in Drinking Water Treatment
Emission Monitoring
Pulp / Paper Industry
Waste Water Treatment
Ozone in Ultra Pure Water
Lowest measuring range 0-0.1 mg/l
Highest measuring range 0-200 mg/l
Applications in Chemical / Pharmaceutical Industry
Industry Measuring Point

Brewing Industry

Colour (EBC) after Beer Filtration
Colour (EBC) of Beer on the Blending Unit
Colour (EBC) of Hot Wort before Cooler

Chemical / Pharmaceutical Industry

Application undisclosed
Colour (Hazen) of Maleic Anhydride
Concentration of Chlorine Dioxide (ClO2)
Concentration of Chlorine in Nitrosyl Chloride (NOCl)
Concentration of Copper in Sulphuric Acid (H2SO4)
Concentration of Hypochlorite (ClO-) in Water
Concentration of Rare-Earth
Ozone in Ultra Pure Water

Drinking Water Treatment

Concentration of Chlorine Dioxide (ClO2)
DOC (UV Absorption) after Activated Carbon Filtration
DOC (UV Absorption) in Raw Water
DOC (UV Absorption) in Raw Water

Emission Monitoring

DOC (UV Absorption) in Raw Water

Food Industry

Concentration of Coffee in Condensate
Concentration of Flavonoids in Lemon Juice

Machine and Metal Industry

Concentration of Copper in Sulphuric Acid (H2SO4)
Concentration of SurTec 650 (chromitAL) in Demineralized Water

Petrochemical Industry / Refinery

Colour (Hazen) of Maleic Anhydride
Concentration of Copper in Sulphuric Acid (H2SO4)

Pulp / Paper Industry

Concentration of Chlorine Dioxide (ClO2)
DOC (UV Absorption) in Raw Water

Spirits Industry

Colour of Spirits at the Blending Unit

Sugar Industry

Colour (ICUMSA) of Liquid Sugar

Waste Water Treatment

DOC (UV Absorption) in Raw Water