GPR-2600 SERIES % OXYGEN ANALYZER

 

Features an advanced galvanic oxygen sensor capable of measuring oxygen concentrations ranging from 0.05% to 100% over a 36 month operating life with no periodic maintenance required, menu driven controls, integral flow meter, and, a full range of analog outputs. Optional temperature control enhances stability.

 

 

Accuracy: < +1% of FS range under constant conditions Analysis: 0-1%, 0-5%, 0-10%, 0-25% FS ranges;
auto-ranging or manually lock on single range
 Application:  Oxygen analysis from 0.05% to 100% in inert, helium, hydrogen, mixed and acid (CO2) gas streams  Approvals:  CE
Area Classification: General purpose Alarms: 2 adjustable form C relay contacts non-latching; “weak sensor” indicator; power failure; system failure
 Calibration:  Ambient or compressed air on0-25% range; or certified
span gas of O2balance N2approximating 80%of analysis
range or one range higher than analysis range
 Compensation:  Barometric pressure and temperature; optional temperature
controlled heated sample system
 Connections:  1/4″ compression tube fittings  Controls:  Water resistant keypad; menu driven range selection, calibration, alarm and system functions
 Data Acquisition:  Selectable data point intervals  Display:  Graphical LCD 5 x 2.75; resolution .001%; displays real time
ambient temperature and pressure
 Enclosure:  Painted aluminum 7.5” x 10.8” x 12.25” panel mount  Flow Sensitivity:  None between 1-5 SCFH, 2 SCFH recommended
 Linearity:  > .995 over all ranges  Pressure:  Inlet – regulate to 5-30 psig, max 100 psig;
vent – atmospheric
 Power:  Universal; specify 110 or 220 VAC with optional heater  Range ID:  1-5V output with 1V representing the lowest range
 Recovery Time:  Air to < 0.1% in < 30 seconds on N2purge  Response Time:  90% of final FS reading < 10 seconds
 Sample System:  Wetted parts: corrosion resistant materials, stainless steel
optional, includes flow meter with integral metering valve
 Sensitivity:  < 0.5% of FS range
 Sensor Model:  GPR-11-32-4 – requires no maintenance  Sensor Life:  36 months inair at 25ºC and 1 atm
 Signal Output:  4-20mA isolated and 0-1V  Temp. Range:  -10º to 45ºC (GPR sensor); -20º to 45ºC (XLT sensor)
 Warranty:  12 months analyzer; 12 months sensor