GPR-1900 SERIES PPM OXYGEN ANALYZER

Features a compact size that lends itself to being integrated into an existing system or custom factory built analyzer, an advanced galvanic trace PPM oxygen sensor capable of precisely measuring oxygen concentrations ranging from 100 PPB to 1% over a 24 month operating life with no periodic maintenance required, menu driven controls, two (2) field selectable alarm set points, and, a full range of analog outputs.

 

Accuracy: < 1% of FS range under constant conditions Analysis: 0-10 ppm, 0-100, 0-1000, 0-1%, 0-25% FS ranges;
auto-ranging or manually lock on single range
Application: Oxygen analysis from 100 ppb to 1% in inert, helium,
hydrogen, mixed and acid (CO2) gas streams
Approvals: CE
Area Classification: General purpose Alarms: 2 adjustable form C relay contacts non-latching; “weak
sensor” indicator; power failure; system failure
Calibration: Certified gas of O2balance N2approximating 80% of
range above analysis range recommended for optimum
Compensation: Barometric pressure and temperature
Connections: 1/8″ compression tube fittings Controls: Water resistant keypad; menu driven range selection,
calibration, alarm and system functions
Display: Graphical LCD 5 x 2.75; resolution .001 ppm; displays real
time ambient temperature and pressure
Enclosure: Painted aluminum 7” x 4” x 4.5” panel mount
Flow Sensitivity: None between 1-3 SCFH, 2 SCFH recommended Linearity: > .995 over all ranges
Pressure: Inlet – regulate to 5-30 psig, max 100 psig;
vent – atmospheric
Power: Universal 100/120/220/250 VAC or 9-28 VDC
Recovery Time: 60 seconds in air to < 10 ppm in < 1 hr on N2purge Response Time: 90% of final FS reading < 10 seconds
Sample System: None Sensitivity: < 0.5% of FS range
Sensor Model: GPR-12-333 – requires no maintenance Sensor Life: 24 months at 25ºC and 1 atm; average O2< 1000 ppm
Signal Output: 4-20mA isolated Temp. Range: 5º to 45ºC
Warranty: 12 months analyzer; 12 months sensor Wetted Parts: Stainless steel